Human clinical
MSK & Extremity MRI
Weight-bearing imaging for orthopedics, sports medicine and joint care.

- 0.25T
- Field strength
- Modular
- Shielded enclosure
- Single-phase
- Power
- AI
- Reconstruction
Overview
What it is for.
- Standing / seated scanning close to the real load-bearing state
- Reduced metal-artifact interference — usable for post-operative follow-up
- Small modular enclosure — about 12 m² — plus single-phase power
- 0.25T C-type open permanent magnet — 24 mT/m, 80 mT/m/ms, under 5 kW
- Three dedicated coils: knee/ankle, wrist/elbow and shoulder

Deployment
Where it can go.
In-clinicMobile & trailerUpper floors
Specifications
Technical data.
Figures confirmed by the engineering team. Items marked TBC are pending.
| Parameter | PANTORIS Formula III |
|---|---|
| Field strength | 0.25T |
| System weight | 2.9 t |
| Magnet type | Permanent magnet — C-type open, passive and active shimming |
| Bore / patient space | 240 mm gantry opening |
| Scan positions | Standing, seated and supine |
| Shielding requirement | Modular enclosure — about 12 m², confirmed per site |
| Power supply | Single-phase 220/110 Vac — under 5 kW maximum |
| Floor loading | 2.9 t system weight; no reinforcement in most cases |
| Trailer / mobile compatible | Yes |
| Packed dimensions & weight | TBC |
| Operating temperature & humidity | TBC |
| Regulatory status | Cleared in China and the United States |