Veterinary
Veterinary MRI
Equine and small-animal imaging, from the hoof to the head.

- 0.3T
- Field strength
- Modular
- Shielded enclosure
- Single-phase
- Power
- AI
- Reconstruction
Overview
What it is for.
- Standing sedation scanning for horses — avoids the risk of general anaesthesia
- Dedicated hoof and fetlock coils for the distal limb, where X-ray and ultrasound fall short
- Two configurations: 0.3T U-shape with a 330 mm gap, or 0.25T double-column with 300 mm
- Covers animals up to 200 kg, FOV 100–300 mm, DICOM 3.0 with PACS and patient worklist

Deployment
Where it can go.
Equine clinicMobile & trailer
Specifications
Technical data.
Figures confirmed by the engineering team. Items marked TBC are pending.
| Parameter | PANTORIS Veterinary Series |
|---|---|
| Field strength | 0.3T |
| System weight | 4.5 t (0.3T configuration) · 2.5 t (0.25T configuration) |
| Magnet type | Permanent magnet — U-shape (0.3T) or double-column open (0.25T) |
| Bore / patient space | 330 mm gap (0.3T) · 300 mm gap (0.25T) |
| Scan positions | Standing or recumbent, depending on species and coil |
| Shielding requirement | Modular enclosure — confirmed per site |
| Power supply | 110 / 220 V — 3 kW RF (0.3T) · 2 kW RF (0.25T) |
| Floor loading | 2.5–4.5 t system weight; installs on upper floors |
| Trailer / mobile compatible | Yes |
| Packed dimensions & weight | TBC |
| Operating temperature & humidity | TBC |
| Regulatory status | No medical device registration required for veterinary use |
Gallery
Images.



Talk to us about your site.
Tell us where the system needs to go and what you need to image. We will tell you what is possible.